USER MANUAL
Technical data
Ow-RX
(Rev. 6)
28
X-ray tube characteristics
Manufacturer
Toshiba Japan
Type
D-045
Inherent filtration
at least 1 mm Al
Focus size
0.4 (IEC 60336)
Anode tilt angle
12.5°
Anode material
Tungsten
Nominal max voltage
70 kV
Filament max voltage
3.1 V
Filament max current
3 A
Anode thermal capacity
4.3 kJ
Anode cooling capacity (max)
100 W
Environmental conditions
Operating temperature range
+10°C
÷
+40°C
Relative working humidity (RH) range
30%
÷
75%
Operating atmospheric pressure range
700
÷
1060 hPa
Temperature range for transport and
storing
-20°C
÷
+70°C
Humidity range for transport and
storing
<95 % non-condensing
Minimum atmospheric pressure for
storing and transport
630hPa
Weight of the unit and the removable parts
Gross weight including packaging
30 kg
Net weight of the unit in the standard
configuration
23 kg
Extension arm 60 cm (standard)
2.9 kg
Extension arm 80 cm
3.5 kg
Extension arm 30 cm
1.9 kg
Scissors arm with tube-head support
10 kg
Timer + wall support
5.05 kg
Net weight of the mobile stand
31 kg
Timer + mobile stand support
4.7 kg
Tube-head
5 kg
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...