Immunity test
Test level
IEC 60601-1-2
Compliance
level
EMC environment of use
The RF portable and mobile
communications units should not be
used closer to any part of the Ow-RX,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Radiated RF
IEC 61000-4-3
3 V/m
from 80 MHz to
2.5 GHz
3 V/m
d = 1.2 x
√
P
from 80 MHz to
800 MHz
d = 2.3 x
√
P
from 800 MHz to
2.5 GHz
Conducted RF
IEC 61000-4-6
3 V
from 150 kHz to
80 MHz
3 V
d = 1.2 x
√
P
where "
P
" is the maximum rated
output of the transmitter in watts (W)
according to the transmitter
manufacturer and "
d
" is the
recommended separation distance in
meters (m).
The field strength of the fixed RF
transmitters, determined by an on-site
electromagnetic survey, should be
lower than the compliance level in
each frequency range.
Interference may be verified near
devices marked with the following
symbol: