USER MANUAL
System use
(Rev. 6)
Ow-RX
55
8.6
Exposure with the supplied X-ray button
1.
Using the main keypad, select the exposure time as described in
paragraph 8.4, based on the selected mode.
2.
Move away the distance permitted by the X-ray button cable in the
opposite direction of the X-ray beam.
3.
Press the X-ray emission button and hold it down during the
exposure.
4.
A yellow light and an acoustic signal indicate the start of exposure.
WARNING:
•
The X-ray emission button is a “dead-man” control; therefore it must
be held down during the entire exposure.
•
If the button is released before the end of the exposure, the emission
is automatically stopped; this situation is shown on the display by
the message “
E13
”, if the button was released during pre-heating and
“
E12
” if the button was released during the emission of X-rays.
This message will remain displayed until the “Increase” key
is pressed.
5.
At the end of exposure, the system starts the tube-head cooling cycle
(30 times the exposure time). The time until the end of the pause is
shown on the display.
NOTE:
For all exposure times shorter than 0.20s, the cooling pause is constant
and equal to 6s.
6.
If the X-ray button was already pressed at the end of the cooling
pause, the exposure will be inhibited and the error “
E11
” will be
displayed.
NOTE:
All the system statuses are shown on the display during the exposure:
preheating, X-ray emission and the cooling pause.
Summary of Contents for Ow-RX
Page 1: ...Version July 2019 Rev 7 Ow RX User Manual 0051...
Page 2: ......
Page 4: ...USER MANUAL Revision history Ow RX Rev 6 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 22: ...USER MANUAL Description Ow RX Rev 6 16 5 DESCRIPTION 5 1 Identification plates 5 1 2 6 4 3...
Page 73: ...USER MANUAL Maintenance Rev 6 Ow RX 67...
Page 74: ...Cod 6959902203_Rev 6 0051...