System use
Ow-RX
(Rev. 6)
48
8.4
Preparing the tube-head
1.
Position the tube-head with an angle suitable for the exposure and
positioning required (see Figure 12, Figure 13, Figure 14, Figure 15).
2.
Introduce the image receptor in the patient’s mouth according to
the selected technique (bisector or parallel). In this regard, see
paragraph 8.5.
3.
Move the tube-head cone towards the patient and direct it exactly
towards the tooth to be X-rayed, referring to the following figures.
NOTE:
If you want to use the rectangular collimator, apply it to the end of the
tube-head cone, positioning it as needed.
NOTE:
If you want to use the 30 cm extension cone, apply it to the end of the
tube-head’s 20 cm cone.
WARNING:
If the 30 cm extension cone is applied, the pre-set exposure times will be
automatically doubled in order to obtain the same radiographic result.