8.7.1
Indication of the battery charge status and replacement.
The wireless X-ray button diagnoses the status of battery.
If the battery level is lower than 2.7V, the remote control informs the user
with 2 sequences of 5 very high flashes.
NOTE:
In this situation, a few exposures can be made, but the batteries should
be replaced as soon as possible.
Proceed as follows to replace the batteries:
•
unscrew the two screws located on the back of the button
•
open the two half-shells, keeping the green button facing upwards
and paying attention to the electronics located inside
•
replace the batteries respecting the indicated polarities
•
reclose the two half-shells and tighten the screws.
If the charge level is lower than 2.4V, the LED on the remote control will
perform 4 sequences of 5 very high flashes and exposures cannot be
made.
8.7.2
Matching procedure between the remote control and timer
The wireless X-ray button will only work with the timer with which it was
combined.
If for some reason the two devices must be combined again, proceed as
described in the Service Manual.