Otoport OAE+ABR
User Manual for Otoport DP+TE model
67
OAE test procedure
The following table describes which highlighted message will appear if
more than one condition is met.
Note that highlighted messages also appear during the test.
Consistent
Stimulus Out
Stimulus
Highlighted
High Noise
of Range
Ringing
Message
No
No
No
Checkfit
Yes
No
No
Noisy
No
Yes
No
Check Probe Fit
No
No
Yes
Ringing
Yes
Yes
No
Check Probe Fit
Yes
No
Yes
Ringing
No
Yes
Yes
Check Probe Fit
Yes
Yes
Yes
Check Probe Fit