160
Quality checks
CHAPTER SEVENTEEN
Quality checks
Replacement probes supplied should be tested with your
Otoport system and new Probe Test results saved as a baseline
reference.
On selection of
Probe Test
the message
Place Probe into Otodynamics
Test Cavity. Press OK to begin test.
will appear on screen.
If two probes are connected, then the probe test will run automatically on
both, so it is necessary to insert both into test cavities.
Remove the tip from the probe and place the cavity on a flat surface. Insert
the probe into the test cavity at a 90 degree angle to the top of the cavity,
between the screws, as shown below left. Press the probe firmly into the
cavity until the shoulder of the probe touches the top of the cavity. When
released, the probe will rise a little to its natural position and the shoulder
may no longer touch the cavity. Inserting the probe at the wrong angle
or with the probe head over one of the screws may result in incorrect test
results.
Select
OK
to begin the
Probe Test
or
Cancel
to return to the
Probe Menu
screen.