Otoport OAE+ABR
User Manual for Otoport DP+TE model
189
Equipment troubleshooting
22.5
Hardware fault messages
The Otoport performs a series of hardware tests when it is first turned
on. In the event of a fault being detected the following message will be
displayed:
The error code number displayed indicates the type of error detected. You
should make a note of this error number. The Otoport should then be
turned on and off a number of times to ensure that the error doesn’t recur.
If you receive the hardware fault message again, contact your distributor or
Otodynamics for support.
Error 2
An Error 2 message indicates that excessive noise was detected during
start up. The noise may have been detected through the probe, if it was
connected. Noise detected through the probe does not indicate a fault. If
this message is displayed, turn the unit off, disconnect the probe and then
turn it back on. If the message is consistently displayed with the probe
disconnected, then contact Otodynamics.