Otoport OAE+ABR
User Manual for Otoport DP+TE model
179
Power
21.3
Standby
To save power, the Otoport will go into standby mode after 1.5 minutes of
inactivity.
The Otoport will not go into standby if a test is being performed.
To resume from standby, press any button on the keypad. The Otoport will
wake up and return to the previous screen displayed.
If the Otoport is left for 20 minutes in standby it will turn off. An audible
beep will be emitted from the device for a period of 10 seconds to alert the
user prior to the automatic shut down.
Notes:
Following an OAE recording, always save test data, as data that
has not been saved prior to auto switch off will be lost.
Over time batteries will wear and lose their capacity, resulting in
quicker discharge. The batteries may therefore need replacing
around every 4 years of use.