Otoport OAE+ABR
User Manual for Otoport DP+TE model
123
Configuration
The fixed and configurable test parameters are summarised in the following
tables:
Configurable test parameters
Name
Any 2
Any 3
Any 4
Wide
Narrow
Min pass
bands
2
3
4
None
None
Min total (wide
band) SNR
None
None
None
6dB
6dB
Min SNR per
pass band
6dB
6dB
6dB
None
None
Filter (Hz)
1189-4757 1189-4757 1189-4757 1189-4757 1600-3200
Response
Window (ms)
2-5-12.5
2-5-12.5
2-5-12.5
2-5-12.5
4-10
Fixed test parameters
Min sweeps (NLo)
30
Min total (wide band) OAE signal
0dBSPL
Min band signal
-5dBSPL
Stimulus level
84 +/- 1dB pe
Max test time
300 seconds
Max low noise (NLo)
260 sweeps of 16 stimulus presentations
sweeps averaged
Ring reject
-20dB
Max ratio of stimulus peak to the level 3ms after stimulus presentation.
Stimulus ring can hamper accurate OAE measurement
Input filter
1600-3200Hz - narrow mode only
or 1189Hz - 4757Hz
The attenuation at these frequencies is 3dB. Attenuation increases by 80dB/decade
below and 40dB/decade above these frequencies
The default setting is
Any 2 Bands
. However, it is recommended that you
carefully choose pass criteria which suit your program requirements.