7.3
Facility Power Connection
Although the shore power connection in not an actual physical feature of the mobile unit, it is an
integral part of the daily operations.
Circuit Breaker
Receptacle
Manufacturer: Facility
provided
Manufacturer: Russellstoll
Ampere Rating: 200 A disconnect
Model: DF 2504 FRAB 0
Ampere
Rating: 200
A
Figure 35: Shore Power Connection
Oshkosh Specialty Vehicles
Connector:
The plug that is provided by Oshkosh Specialty Vehicles
for connection to the shore power receptacle.
Connector Lock Ring:
Secures the connections.
Power Cable:
The cable that runs between the shore power connections
and the 480V AC electrical panel.
Shore Power Disconnect:
The shore power disconnect terminates the power to the
receptacle. This must be in the “OFF” position when
connecting to the receptacle.
Shore Power Receptacle Outlet:
The receptacle outlet that the shore facility has installed
for use with the Oshkosh Specialty Vehicles connector
and power cable.
Shore Power Unit:
The complete shore power assembly.