Section 4: Safety Systems
This safety section contains important information about the safety systems that have been built
into the mobile unit to protect all personnel and equipment. Before attempting to service the mobile
unit, read this safety section as well as all other safety sections found in applicable manufacturers’
manuals in the component literature binder.
Figure 21: Interlocks
4.1
Door Interlock System
Scan Room and Magnet Room are shielded from radio frequency interference. If doors that lead to
these rooms are opened, radio frequencies can interfere with the scan image. A door interlock
system has been incorporated into the mobile unit to ensure that proper scanning can take place.
This system provides a constant monitoring of the doors that have the interlock system. If one of
these doors is opened, the technician will see a notification appear at the console stating that an
RF door has been opened. All doors must be closed for scanning to take place. If a door is opened
during a scan, scanning operations will be stopped. The door interlocks can be found at the
following locations.
•
On the interior door that leads into Scan Room.
•
On the exterior door that leads into Magnet Room.