5.11 Switch from Generator Power to Shore Power
When switching from generator power to shore power the “480V AC FAULT”
may illuminate and flicker. If the “480V AC FAULT” stays illuminated,
reconnect to generator power and call a certified electrician before attempting
to reconnect to shore power.
When turning the power selector switch for the incoming power, from one
position to another, the selector must be paused for a minimum of five
seconds, in the “OFF” position between selections. Failure to do so can result
in damage to the equipment.
1. Open the left Side Center Lower Compartment door to access the power source controls.
2. Turn the unit power selector from “Generator” to “Line”. Be certain to pause for five
seconds in between selections.
3. The two indicators on this panel will illuminate according to the power that is received.
a. If the supply of power is good, the green indicator labeled “480V AC OKAY” will
illuminate.
b. If the supply of power is experiencing problems, such as being out of phase, the
red indicator labeled “480V AC FAULT” will illuminate.
4. Since minor fluctuations in power are common, the “480V AC FAULT” may briefly flicker.
This is to be expected. If the “480V AC FAULT” indicator light stays illuminated, switch
back to generator power and contact a qualified electrician before operating the medical
system.
5. If the power supply is good, the selector for the generator can be moved to the “STOP”
position. Once this has been done, the generator will enter into a cooling phase and
continue running for approximately five minutes. Do not continue to move the selector
switch during this time, the generator will shut down automatically when this stage has
completed.
6. Turn the Generator selector switch to the “OFF” position. Be certain to pause for five
seconds in between selections.
5.12 Connect the Phone and Data Lines
The phone and data lines can be found in the underbody storage compartments. Both the phone
and data lines can now be connected from the outlets located in the underbody compartments to
the receptacles located at the shore site.
The phone lines make use of Hubbell all weather connections. The data lines make use of CAT-5E
cable and RJ-45 connections. Refer to
Figure 15: Phone & Data Line Connections