Version 4.2: Aug 26, 2015
17
b. Room Ventilation:
The room ventilation of the laboratory should preferably be a ventilation system that complies with the national
requirements for the application.
•
The inlet air and exhaust air openings of the room ventilation must be located so that drafts are prevented
from impairing the function of the MARS air system.
c. Installation test
Do not operate the equipment before the installation test has been completed.
The installation inspection of the workstation must be conducted by a qualified technician or a certified company. It
is important to validate if the workstation functions listed below were checked and if the results are within the safety
value tolerances specified in the manufacturing test report included in the folder.
Down flow velocity test
Inflow velocity test
HEPA Filter Leak Test
Particle count test
Temperature calibration of the heat surface and the heated glass. (MARS IVF only)
A repeat test must also be performed after repairs to the device or after moving the device
more than 10 cm (4.9 inch).
The operator must request a written test report from the authorized test service.
Safety warranty: The operational safety of the device, particularly the samples, is guaranteed
only if all safety functions of the device have been tested and approved. ORIGIO will not warrant
the operational safety if the device is operated without performance of the required installation
test or if the installation test and repeat test are not performed by adequately trained and
authorized personnel.
Device Hygiene: The installation test does not include any decontamination measures. For
operation in the sample chamber of the workstation, the sample chamber and the accessories
required for the work process must be disinfected and cleaned in accordance with the hygiene
guidelines set forth for the application.
h. Accessories
CAUTION:
The installation of the workstation must be completed by a certificated or technically
trained engineer approved by ORIGIO or our authorized distribution network
There are several accessories that can be either integrated or added to the MARS workstation. These concerns:
Light source: TLB 7000
Heated glass stage: HG37
Humidifier
Integrated Stereomicroscope(s)
Camera(s) for the stereomicroscope(s)