58
Version 4.2: Aug 26, 2015
FLOW ALARM 1 with HIGH and a set of values will be displayed (e.g. 0285)
Press on the arrow down until the alarms starts. By pushing the arrow the
numbers will decrease by units.
By maintaining a constant pressure on the arrow, the numbers will decrease by
decimals
Press on the arrow up one unit at a time and wait for 2 to 3 seconds to see if the
alarm stops.
Repeat the operation until the alarm stops.
When the alarm has stopped, read the value displayed: e.g. 0225. Add 50 to the
number: 0225 + 50 = 0275. Press on the arrow up until you reach this number.
Press on the “ENTER” button.
FLOW ALARM 1 with LOW and a set of values will be displayed (e.g. 0175)
Press on the arrow up until the alarms starts. By pushing the arrow the numbers
will decrease by units.
By maintaining a constant pressure on the arrow, the numbers will decrease by
decimals.
Press on the arrow down one unit at a time and wait for 2 to 3 seconds to see if
the alarm stops.
Repeat the operation until the alarm stops.
When the alarm has stopped, read the value displayed: e.g. 0125. Add 50 to the
number: 0125 + 50 = 0175. Press on the arrow down until you reach this
number.
Press on the “ENTER” button.
FLOW ALARM 1 with ALARM NORMAL will be displayed.
Press on the “ENTER” button and navigate in the menu until your reach
SERVICE LOGOUT and press “ENTER” you will now reach the standard menu.