64
Version 4.2: Aug 26, 2015
XI.
TECHNICAL SPECIFICATIONS
This manual covers the following MARS Models:
a. MARS 900 IVF or LAF
IVF model is illustrated
MARS 900 IVF
Dimensions (W x D x H)
1003 X 797 x 2050 mm
Worksapce (WxD XH)
900 x 593 x 698 mm
Table plate standard height
80 – 85 cm (option 75-80 cm or 90 -95 cm)
Air velocity, vertical flow
0,15 m/s (adjustable 0.01 -0.70 m/s)
Air Velocity, deviation
+/- 10%
Noise Level, ISO 6081
<52 dB(A)
Light intensity variable
0 – 2000 Lux
HEPA Filters, EN 1822
Efficiency is 99.999% against 0.3
µ
m particle H-14 size
Power consumption
125 W
Power consumption normal operation
LAF: 270 W
IVF: 660 W (heat plate warm up)
Max power consumption outlets (total)
400 W
Main overcurrent protection fuses
Workstation
Rating
LAF
IVF
230V
3.15 A
6.3 A
115 V
6.3 A
8 A
Outlet overcurrent protection fuse
Workstation
Rating
LAF
IVF
230V
3.15 A
3.15 A
115 V
4 A
4 A
Voltage / Frequency
220-240V / 50-60 Hz or 110 -120V / 50 -60 Hz
Window material ( side/Front)
Hardened / laminated safety glass
Cabinet material / Work surface
Polyester coated steel /AISI 304 stainless steel
Ingress Protection
IP50
Pollution Degree
2
Intended use
Indoor Use Only
External dimensions packed (D x W x H)
1516x916x1559 mm
Shipping volume
2,17 m³
Net weight / Gross Weight
225 Kg / 315 Kg