44
NOTE: when choosing this ramping option, make sure the
temperature
is a negative value
, otherwise the bath temperature
will increase instead of decrease.
•
Condition: determines the condition under which the above set bath
temperature is active. Pressing the
🔄
symbol gives a few different
condition options:
o
ºC ST: this is the sample temperature (for instance: -20 ºC) until
which the above set bath temperature is active. When this sample
temperature is reached, the next bath step will be activated.
o
sec.: this is the amount of seconds the bath temperature will be
active until the next bath step will be activated.
o
min.: this is the amount of minutes the bath temperature will be
active until the next bath step will be activated.
•
hours.: this is the numbers of hours the bath temperature will be active
until the next bath step will be activated.
•
END: The set bath temperature will be active until the end of the test.
End of test conditions are set in “End conditions”)
After entering the settings, press DONE to create the step and exit the menu.
For multiple steps, repeat the above procedure until desired cooing profile is
complete.
🔄
Auto number = 0:
If this line still appears in your latest version of the software: please ignore it as
it is a leftover line from an old feature that is not present nor relevant anymore.