15
TECHNICAL SPECIFICATIONS:
AirSTAR CFPP Analyzer
AirSTAR CPPP Analyzer
Methods
EN 116, IP 309, ASTM D6371, JIS K
2288, EN 16329.
Cloud Point: ASTM D2500, D5771, EN 219, ISO 3015, JIS
K2269
Pour Point: ASTM D97, D5853, D6892, EN 15, ISO 3016,
JIS K2269m, Chinese GB/T3535
Sample
detection
3 laser beam detection barriers
Camera, high definition. Detection of entire bottom area
and throughout entire sample.
Calibration
Extensive calibration routines for bath jacket temperature (and vacuum measurement for
CFPP).
Temperature probe offset table. Programmable frequency of calibration notifications
Bath temp
range:
+65°C down to -105°C (-175ºF to 95°F)
Vacuum
Microprocessor controlled; constant and
dynamic profiles customizable.
Not applicable
Lift
mechanism
Not applicable
Microprocessor controlled stepper motor
Test/Aspirati
on Settings
Test can be started at programmable temperatures. Option for customized time and
aspiration/measuring intervals.
Cooling
profiles
Stepped and linear cooling profiles customizable (from 1 to 60°C/h). Unlimited storage capacity.
Temperatur
e
Measureme
nt
°C or °F; PT 100 class A probe, automatic probe ID detection with built-in calibration data storage.
Calibration certificate standard supplied. Metal probe for CPPP Head. In-glass probe for CFPP Head
(metal probe optionally available)
Cleaning
Quick programmable cleaning cycles, in
addition to required cleaning of filter
according to method specifications.
Cleaning is executed in Cleaning-Docking
station.
Manual cleaning of jar and probe.
Voltage
100–240 VAC, 50/60 Hz
Power
150 W
Dimensions
16cm x 59cm x 35cm (W x D x H)
16cm x 59cm x 49cm (W x D x H)
Weight
25 kg