Technical Data
27
-
2.11 Measurement specifications [G5 models]
G5L
G5H
G5HF
Temperature range (scalable)
100...1200 °C
250...1650 °C
200...1650 °C
Spectral range
5,0 µm
Optical resolution
45:1
70:1
45:1
System accuracy
1), 2), 3)
(at ambient
temperature 23
5 °C)
±1.5 °C or ±1 %
Repeatability
1)
(at ambient temperature
23
5 °C)
±0.5 °C or ±0.5 %
Temperature resolution (NETD)
0,1 K
Response time (90 % signal)
120 ms
80 ms
10 ms
Emissivity/ Gain
0.100...1.100 (adjustable via programming keys or software)
Transmissivity
0.100...1.000 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
Whichever is greater,
2)
= 1/ Response time 1s,
3)
Accuracy for thermocouple output: ±2.5 °C or ±1
%,
whichever is greater