Technical Data
25
-
2.9 Measurement specifications [3M/ MT/ F2 models]
3MH3
4)
MT
MTH
F2
Temperature range (scalable)
250...1800 °C
200...1450 °C
400...1650 °C
200...1450 °C
Spectral range
2.3 µm
3.9 µm
4.24 µm
Optical resolution
300:1
45:1
System accuracy
1), 6)
(at ambient
temperature 23
5 °C)
±(0.3 % T
Meas
+2 °C)
±1 %
2)
Repeatability(at ambient temperature
23
5 °C)
±(0.1 % T
Meas
+1 °C)
±0.5 % or 0.5 °C
2)
Temperature resolution (NETD)
0.1 K
Response time (90 % signal)
1 ms
3)
10 ms
Emissivity/ Gain
0.100...1.100 (adjustable via programming keys or software)
Transmissivity
0.100...1.000 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
= 1/ Response time 1s,
2)
at object temperatures >300 °C,
3)
with dynamic adaptation at low signal levels,
4)
T
Object
> T
Head
+25 °C,
5)
Specification valid at T
Object
≥ start of measurement range + 50°C,
6)
Accuracy for thermocouple output: ±2.5 °C or ±1
%
Summary of Contents for CTlaser 05M
Page 1: ...Operator s Manual optris CTlaser LT LTF 05M 1M 2M 3M MT F2 F6 G5 P7 Infrared thermometer...
Page 8: ...8 Appendix C Smart Averaging 96 Appendix D Declaration of Conformity 97...
Page 44: ...44 Figure 2 Mounting bracket adjustable in one axis Order No ACCTLFB standard scope of supply...
Page 45: ...Mechanical Installation 45 Figure 3 Electronic box...
Page 97: ...Appendix D Declaration of Conformity 97 Appendix D Declaration of Conformity...
Page 98: ...optris CTlaser E2018 12 A...