68
-
4.7.4
Functional inputs
The three functional inputs F1
– F3 can be programmed with the CompactConnect software, only.
F1 (digital):
trigger (a 0 V level on F1 resets the hold functions)
F2 (analog):
external emissivity adjustment
[0–10 V: 0 V ►
= 0.1; 9 V ►
= 1; 10 V ►
= 1,1]
F3 (analog):
external compensation of ambient temperature/ the range is scalable via software
[0–10 V ► -40–900 °C/ pre-set range: -20–200 °C]
F1-F3 (digital):
emissivity (digital choice via table)
A non-connected input represents:
F1 = High | F2, F3 = Low
[High level: ≥ +3 V…+36 V | Low level: ≤ +0,4 V…–36 V]