2.6 Measurement specifications [1M models]
1ML
1MH
1MH1
Temperature range (scalable)
485...1050 °C
650...1800 °C
800...2200 °C
Spectral range
1,0 µm
Optical resolution
150:1
300:1
System accuracy
1), 3)
(at ambient temperature
23
5 °C)
±(0.3 % T
Meas
+2 °C)
Repeatability (at ambient temperature 23
5 °C)
±(0.1 % T
Meas
+1 °C)
Temperature resolution (NETD)
0.1 K
Response time
2)
(90 % signal)
1 ms
Emissivity/ Gain
0.100...1.100 (adjustable via programming keys or software)
Transmissivity
0.100...1.000 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
ε = 1, Exposure time 1 s,
2)
with dynamic adaptation at low signal levels,
3)
Accuracy for thermocouple output: ±2.5 °C or ±1
%,
whichever is greater