62
-
4.5 Exchange of the sensing head
After exchanging a head the calibration code of the new head must be entered into the
electronics.
After modification of the code a reset is necessary to activate the changes.
[►
The calibration code is fixed on a label on the head. Do not remove this label or note the
code. The code is needed if the electronic must be exchanged.
The sensing head is already connected to the electronics by factory default. Inside a certain model group an
exchange of sensing heads and electronics is possible.
Entering of the calibration code
Every head has a specific calibration code, which is printed on the head. For a correct temperature
measurement and functionality of the sensor this calibration code must be stored into the electronic box. The
calibration code consists of five blocks with 4 characters each.
Example:
EKJ0
– 0OUD – 0A1B – A17U – 93OZ
block 1 block 2 block 3 block 4 block 5