2.8 Measurement specifications [3M models]
3ML
1)
3MH
1)
3MH1
2)
3MH2
2)
Temperature range (scalable)
50...400 °C
100...600 °C
150...1000 °C
200...1500 °C
Spectral range
2.3 µm
Optical resolution
60:1
100:1
300:1
System accuracy
3)5)
(at ambient
temperature 23
5 °C)
±(0.3 % T
Meas
+2 °C)
Repeatability
(at ambient temperature
23
5 °C)
±(0.1 % T
Meas
+1 °C)
Temperature resolution (NETD)
0.1 K
Response time
4)
(90 % signal)
1 ms
Emissivity/ Gain
0.100...1.100 (adjustable via programming keys or software)
Transmissivity
0.100...1.000 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
T
Object
> T
Head
+25 °C,
2)
Specification valid at T
Object
≥ start of measurement range + 50°C,
3)
= 1/ Response time 1s,
4)
with dynamic
adaptation at low signal levels,
5)
Accuracy for thermocouple output: ±2.5 °C or ±1
%,
whichever is greater