Technical Data
23
-
2.7 Measurement specifications [2M models]
2ML
2MH
2MH1
Temperature range (scalable)
250...800 °C
385...1600 °C
490...2000 °C
Spectral range
1.6 µm
Optical resolution
150:1
300:1
System accuracy
1)3)
(at ambient temperature
23
5 °C)
±(0.3 % T
Meas
+2 °C)
Repeatability (at ambient temperature
23
5 °C)
±(0.1 % T
Meas
+1 °C)
Temperature resolution (NETD)
0.1 K
Response time
2)
(90 % signal)
1 ms
Emissivity/ Gain
0.100...1.100 (adjustable via programming keys or software)
Transmissivity
0.100...1.000 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
= 1, Exposure time 1 s,
2)
with dynamic adaptation at low signal levels,
3)
Accuracy for thermocouple output: ±2.5 °C or ±1
%,
whichever is greater
Summary of Contents for CTlaser 05M
Page 1: ...Operator s Manual optris CTlaser LT LTF 05M 1M 2M 3M MT F2 F6 G5 P7 Infrared thermometer...
Page 8: ...8 Appendix C Smart Averaging 96 Appendix D Declaration of Conformity 97...
Page 44: ...44 Figure 2 Mounting bracket adjustable in one axis Order No ACCTLFB standard scope of supply...
Page 45: ...Mechanical Installation 45 Figure 3 Electronic box...
Page 97: ...Appendix D Declaration of Conformity 97 Appendix D Declaration of Conformity...
Page 98: ...optris CTlaser E2018 12 A...