OPERATING INSTRUCTIONS 

MODEL: 111 

INFRARED THERMOMETER 

WITH LASER SIGHTING 

INTRODUCTION 

This instrument is a portable easy to use 3½digit, com-

pact-sized digital infrared thermometer with laser sighting 

designed for simple one hand operation. Meter comes with 

backlight LCD display, Auto-Hold function and Auto 

Power Off (10 seconds approx.) to extend battery life. 

  

SAFETY INFORMATION 

It is recommended that users read the safety and opera-

tion instructions before using the infrared thermometer. 

 DANGER 

When 

  sign appears, pull the trigger to turn the laser 

beam on and off. 

Be very cautious with the laser beam, not aiming your 

eyes or those of any other person or animal. 

 Do not look directly into the laser light from the opti-

cal system. 

 When measuring the temperature of an object which 

has a mirror finish, be careful not to allow the laser 

light beam to be reflected off the surface into your 

eyes or those of another person. 

  Do not allow the laser light beam to impinge upon any 

gas which can explode. 

CAUTION 

 Do not use the unit near any device which generates 

strong electromagnetic radiation or near a static elec-

trical charge, as these may cause errors. 

  Do not use the unit where it may be exposed to corro-

sive or explosive gases. The unit may be damaged, or 

explosion may occur. 

 Do not keep or use this unit in an environment where 

it will be directly illuminated by sunshine, or where it 

condensation. If you do, it may be deformed, its insu-

lation may be damaged, or it may no longer function 

according to specification. 

 Do not place the meter on or around hot objects 

(70°C/158°F). It may cause damage to the case. 

 If the meter is exposed to significant changes in ambi-

ent temperature (hot to cold or cold to hot). Allow 30 

minutes for temperature stabilization, before taking 

measurement. 

 Condensation may form on the sensor when going 

from a cold to hot environment-wait 10 minutes for 

condensation to dissipate before taking measurements.

 This unit is not constructed to be waterproof or dust

proof, so do not use it in a very dusty environment or 

in one where it will get wet. 

 Do not point the lens at the sun or at any other 

source of strong light. If you do, the sensor may be 

damaged.

 Do not contact the lens against the object whose 

temperature is to be measured, or get it dirty, allow 

it to be scratched, or allow any foreign material to 

adhere to it. Doing so may cause errors.

 Do not touch or hold by the front case. Tempera-

ture reading can be affected by heat from hand. 

 Readings may be affected if the unit is operated 

within a radio frequency electromagnetic field 

strength of approximately 9 volts per meter, but 

the performance of the instrument will not be per-

manently affected.

SPECIFICATIONS 

GENERAL 

Display: 

3½ digit liquid crystal display (LCD) with maxi-

mum reading of 1999. 

Low battery indication: 

The “ ” is displayed when the 

battery voltage drops below the operating level. 

Measurement rate: 

0.25 second per measurement.

Operating Environment: 

32°F to 122°F (0°C to 50°C) at 

< 70% R.H. 

Storage Temperature: 

-4°F to 140°F (-20°C to 60°C), 0 

to 80% R.H. with battery removed from meter. 

Auto power off: 

10 seconds. 

Standby consuming current: 

<5µA. 

Battery: 

Standard 9V battery (NEDA 1604, IEC 6F22 

006P). 

Battery Life: 

9 hours (continuity) typical. 

(contain Laser and Backlit) 

Dimensions: 

148mm(H) x 105mm(W) x 42mm(D). 

Weight: 

approx. 157g. (including battery) 

Laser Specifications 

Laser safety classification of Class 2. 

Wave Length: 

Red (630~670nm). 

Power out: 

<1mW, class 2 laser product. 

ELECTRICAL 

Temperature Range: 

-30°C to 550°C / -22°F to 1022°F. 

Display Resolution: 

0.5/1°C (Auto), 1°F. 

Accuracy: 

  ±(2°C/4°F) for -30°C to 100°C, -22°F to 212°F. 

  ±(2% reading) for 101°C to 550°C, 213°F to 1022°F.

Temperature Coefficient: 

±0.2% of reading or ±0.36°F/0.2°C, whichever is 

greater, change in accuracy per °F/°C change in ambient 

operating temperature above 82.4°F/28°C or below 

64.4°F/18°C. 

Response Time: 

0.25 second 

Spectral Response: 

6 to 14µm 

Adjustable emissivity (

ε

): 

0.1 to 1.0. 

Detection Element: 

Thermopile. 

Optical Lens: 

Fresnel Lens. 

Sighting: 

1-beam laser marker <1mW (class 2). 

Field of View: 

100mmØ at 1000mm (3.9"Ø at 39.0"). 

Spot size increases with distance from the probe tip as shown 

(Spot Diameter measured at 90 % Energy)

DEFINITION















FUNCTION KEY   

SET key 

Press “SET” key goes to the setting mode. Use the same 

key to go to next step, following as the cycle : 

ε →

 ALM 

Hi 

 ALM Lo 

 MAX 

 MIN 

 measuring mode. 

 key 

Press    key, LCD shows “ ” sign. Then pull the trig-

ger to turn on laser pointing function to aim at the measur-

ing point of object. Repeat same steps turns off laser point-

ing function. 

 key 

Use    key to turn on or off the Back-Light function. 

°C/°F key 

Use the °C/°F key to select reading displayed in Cel-

sius(°C) or Fahrenheit(°F). 

APO key 

The meter is default to APO (auto power off) activated 

along with manual mode. Leave the meter without opera-

tion for 10 seconds, power turns off automatically.

Press “APO” key to disable APO function and will go to 

auto continuous mode. Continuous measuring will be 

performed without holding the trigger. Press “APO” key 

again to return to APO activated along with manual mode. 

 and 

 key 

Use 

 or 

 to adjust 

ε

, ALM Hi, and ALM Lo in 

setting mode. Hold the key will speed up.   