background image

MT 700

IB MT 700 EN-PL 3320 - Revision Date: 2020-08-21

Description of this Thermometer

 

Important Safety Instructions

Follow instructions for use. This document provides important 

product operation and safety information regarding this device. 

Please read this document thoroughly before using the device and 

keep for future reference.

This device is only to be used for measuring human body tempera-

ture through oral, rectal or axillary. Do not attempt to take tempera-

tures at other sites, such as in the ear, as it may result in false read-

ings and may lead to injury.

Do not use this device if you think it is damaged or notice anything 

unusual.

We recommend cleaning this device according to the cleaning 

instructions before first use for personal hygiene.

The minimum measurement time until the beep is heard must be 

maintained without exception!

Consider that different measurement locations may require 

continued measuring even after the beep, see section «Measuring 

methods / Normal body temperature».

Do not attempt rectal measurements on persons with rectal disor-

ders. Doing so may aggravate or worsen the disorder.

Ensure that children do not use this device unsupervised; some 

parts are small enough to be swallowed. Be aware of the risk of 

strangulation in case this device is supplied with cables or tubes.

Do not use this device close to strong electromagnetic fields such as 

mobile telephones or radio installations. Keep a minimum distance 

of 3.3 m from such devices when using this device.

Protect the device from impact and dropping!

Avoid bending the thermometer probe more than 45°!

Avoid ambient temperatures above 60 °C. NEVER boil this device!

Use only the commercial disinfectants listed in the section «Cleaning 

and Disinfecting» to clean the device to avoid damage to the device.

We recommend this device is tested for accuracy every two years or 

after mechanical impact (e.g. being dropped). Please contact your 

local Microlife-Service to arrange the test.

WARNING:

 The measurement result given by this device is 

not a diagnosis! Do not rely on the measurement result only.
Batteries and electronic devices must be disposed of in 

accordance with the locally applicable regulations, not with 

domestic waste. 

Read the instructions carefully before using this device.

Type BF applied part

Turning on the Thermometer

To turn on the thermometer, press the ON/OFF button 

1

; a long beep 

signals «thermometer ON». A display test is performed. All segments 

should be displayed.

Then «

-- -

» and a flashing «

°C

» appear at the display field 

2

. The 

thermometer is now ready for use.

Function Test

Correct functioning of the thermometer is tested automatically each 

time it is turned on. If a malfunction is detected (measurement inaccu-

racy), this is indicated by «

ERR

» on the display, and a measurement 

becomes impossible. In this case, the thermometer must be replaced.

Using the Thermometer

Choose the preferred measuring method. When taking a measure-

ment, the current temperature is continuously displayed and the «

°C

» 

symbol flashes. If the beep is heard 10 times and the «

°C

» is no longer 

flashing, this means that the measured increase in temperature is less 

than 0.1 °C in 5 seconds.

To prolong the battery life, turn off the thermometer by briefly pressing 

the ON/OFF button 

1

. Otherwise the thermometer will automatically 

turn off after about 10 minutes.

Reviewing the last Measurement

When turning on the thermometer, the last measurement will be 

displayed with a «

M

» for memory.

1

ON/OFF button

2

Display

3

Battery compartment cover

4

Measuring sensor / measuring tip

5

Cleaning and disinfecting area (thermometer probe only)

Microlife AG, Espenstrasse 139

9443 Widnau / Switzerland

EN

Measuring methods / Normal body temperature

In the armpit (axillary) / 34.7 - 37.3 °C

Wipe the underarm with a dry towel. Place the measuring sensor 

4

 

under the arm into the center of the armpit so the tip is touching the 

skin and position the patient’s arm next to the patient’s body. This 

ensures that the room air does not affect the reading. Because the 

axillary takes more time to reach its stable temperature wait 

at least 

5 minutes

, regardless of the beep sound.

In the mouth (oral) / 35.5 - 37.5 °C

Do not eat or drink anything hot or cold 10 minutes before the 

measurement. The mouth should remain closed up to 2 minutes 

before starting a reading.

Position the thermometer in one of the two pockets under the 

tongue, to the left or right of the root of the tongue. The measuring 

sensor 

4

 must be in good contact with the tissue. Close your mouth 

and breathe evenly through the nose to prevent the measurement 

from being influenced by inhaled/ exhaled air.

If this is not possible due to blocked airways, another method for 

measuring should be used.

Approx. measuring time: 30 seconds!

In the anus (rectal) / 36.6 - 38.0 °C

Carefully insert the measuring sensor 

4

 of the thermometer 2 to 

3 cm into the anal aperture.

The use of a probe cover and the use of a lubricant is recommended.

If you are unsure of this measurement method, you should consult a 

professional for guidance/training.

Approx. measuring time: 30 seconds!

Cleaning and Disinfecting

For disinfection in home use environment, use a 70% Isopropyl 

alcohol swab, or a cotton tissue moistened with 70% Isopropyl alcohol 

to wipe surface pollutants off the thermometer probe (note: consider 

the application and safety instruction of the disinfectant manufacturer). 

Always start wiping from the end of the thermometer probe (approx. at 

the middle of the thermometer) towards the thermometer tip. After-

wards the entire thermometer probe (see number 

5

 in the drawing) 

should be immersed in 70% Isopropyl alcohol for at least 5 minutes 

(max. 24 hours). After immersion, let the disinfectant dry off for 1 

minute before next use. Avoid immersing or wiping the display to 

protect it from fading.
The thermometer is not intended for professional use.

Battery Replacement

When the «

» symbol (upside-down triangle) appears in the display, 

the battery is flat and needs replacing. To replace the battery remove 

the battery compartment cover 

3

 from the thermometer. Insert the 

new battery with the + at the top. Make sure you have a battery of the 

same type to hand. Batteries can be purchased at any electrical store.

Technical Specifications

This device complies with the requirements of the Medical Device 

Directive 93/42/EEC.

Technical alterations reserved.

Guarantee

This device is covered by a 

2 year guarantee

 from the date of 

purchase. During this guarantee period, at our discretion, Microlife will 

repair or replace the defective product free of charge.

Opening or altering the device invalidates the guarantee.

The following items are excluded from the guarantee:

Transport costs and risks of transport.

Damage caused by incorrect application or non-compliance with the 

instructions for use.

Damage caused by leaking batteries.

Damage caused by accident or misuse.

Packaging/storage material and instructions for use.

Regular checks and maintenance (calibration).

Accessories and wearing parts: Battery.

Should guarantee service be required, please contact the dealer from 

where the product was purchased, or your local Microlife service. You 

may contact your local Microlife service through our website: 

www.microlife.com/support

Compensation is limited to the value of the product. The guarantee will 

be granted if the complete product is returned with the original invoice. 

Repair or replacement within guarantee does not prolong or renew the 

guarantee period. The legal claims and rights of consumers are not 

limited by this guarantee. 

Type:

Maximum thermometer

Measurement range:

32.0 °C to 43.9 °C

Temp. < 32.0 °C: display «

-- -

» for low (too low)

Temp. > 43.9 °C: display «

H

» for high (too high)

Measurement accuracy:

± 0.1 °C; 34 °C - 42 °C

± 0.2 °C; 32.0 - 33.9 °C and 42.1 - 43.9 °C

Operating conditions:

10 - 40 °C; 15-95 % relative maximum humidity

Storage conditions:

-25 - +60 °C; 15-95 % relative maximum humidity

Battery:

LR41 (1.5V) / SR41 (1.55V)

Battery lifetime:

approx. 4500 measurements (using a new battery)

IP Class:

IP22

Reference to standards:

EN 12470-3, clinical thermometers;

ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC); 

IEC 60601-1-11

Expected service life:

5 years or 10000 measurements

Reviews: