Installation
37
-
Figure 19: Line scanner settings
Take care to position of the scan line between the
rollers of the conveyor system to avoid possible
reflections.
You can use the
A
utomatic
L
ine
A
djustment (
ALA
)
function. The scan line is automatically set over the
temperature for the entire scan width. There must be a
hot object (e.g., glass) on the rollers in the scan line
field. Detailed instructions for setting up the line
scanning function can be found in the folder
Documentation
/
Manuals
/
PIX Connect-MA-E20xx-xx-
X.pdf
The system is now ready and can be operated.
Additional settings, described in
, are required for the correct and undistorted display of the glass
surfaces and the calculation of the glass surfaces.
Now you can use the system and start your temperature measurement.