Technical Data
19
-
2.3.2
Pyrometer
The following optical chart show the diameter of the measuring spot in dependence on the distance between
measuring object and sensing head. The spot size refers to
90 % of the radiation energy
.
The distance is always measured from the front edge of the sensing head.
As an alternative to the optical diagrams, the
. The app can be downloaded for free from the
Google Play store (see QR code).
D = Distance from front of the sensing
head to the object
S = Spot size
The size of the measuring object and the optical resolution of the infrared thermometer determine the
maximum distance between sensing head and measuring object.
In order to prevent measuring errors the object should fill out the field of view of the optics completely.
Consequently, the spot should at all times have at least
the same size
like the object or should be
smaller
than
that.
Summary of Contents for Bottom-Up GIS 450i G7
Page 1: ...Operators Manual Bottom Up GIS 640i G7 and 450i G7 Glass Inspection System ...
Page 11: ... Technical Data 11 Figure 1 Dimensions mm shutter system ...
Page 12: ... 12 Figure 2 Dimensions mm control cabinet ...
Page 13: ... Technical Data 13 Figure 3 Dimensions mm CTL 4M ...
Page 20: ... 20 CTL 4M FF optics Figure 5 spot size diagram ...
Page 27: ... Installation 27 Figure 10 camera distance ...
Page 44: ... 44 Figure 27 Linescan in PIX Connect software customized Layout ...
Page 52: ... 52 Appendix A Control cabinet Figure 34 Wiring diagram of control cabinet ...
Page 57: ...Bottom Up GIS MA E2022 04 A ...