Keratron
®
Scout Installation and Operating Manual
OPTIKON 2000
6.3.3 INTRA-OPERATORY
CONFIGURATION
For this configuration (refer to Fig.5):
1.
Caution: always block the wheels under the basement of the OR
Trolley Base (18).
.I
2. Turn the OR base switch OFF. Connect the PC to the
OR trolley base
through a USB cable, connect the
OR Footswitch
to the base and turn the
OR base ON.
.I
3. Insert the VK into the staff (22) and connect the connector to it.
Caution:
before inserting the connector in the VK take care that the OR Trolley
Base is OFF.
Tight both the side knobs (then the
Mires Group
(23) can be
tilted by the knobs themselves). Then release the arm (17) by loosing the
knob located on the arm itself.
4. Caution: for a sterile use, it is appropriate to envelope the entire VK
by a disposable Sterile Sleeve. For the proper insertion and usage of
the sleeves, refer to the relevant instruction for use.
5. At this point, the images can be acquired by setting the VK parallel to the
patient and using the keys upside-down on display (21) (with the help of an
assistant, if it is not covered with the
Sterile Sleeve
). To take the image you
can use the footswitch or the OK button at the bottom left, by keeping it
pressed and getting close to the eye until the acquisition is automatically
taken. The last image can be canceled by the CLR button, or it can be
processed by the P key and the test saved by the S key. As an alternative
the acquisition can be activated from the PC keyboard, by pressing
“ctrl”+”D” or “ctrl”+”S”. You can also toggle “Auto-Trigger” by
“ctrl”+”Spacebar” (Caution to avoid harsh movements and to clear bad
shots). To enable P-S-C keys, and for more explanation about “auto-
trigger” function, refer however to Options “Process” and to the ”Help-on
line” directly on your PC.
WARNING
Caution: use only PC having an insulation class II of EN 60950 (shortly represented with the
symbol ), i.e.: compliant EN 60601-1 standard.
Otherwise plug your PC to a mains outlet independent from the OR Trolley. In such a case the
PC (as well as the printer, if it is connected during the operation) must also be kept at a
minimum distance of at least 1mt from the patient just like any non-medical equipment.
Cod. 161204EN
6-6
2008-12-03 Rev.B