Plug the power chord only of a PC having an insulation class II of
EN60950 (shortly represented with the symbol
), i.e.: compliant EN
60601-1 standard and then lock it by screwing the Perspex Plate in its
place
The Mains socket-outlet of the OR trolley must be used only to energize
equipment compliant with EN 60601-1 standard (or class II of EN60950). The
risks that are encountered when equipment, whose compliance with EN 60601-
1 is unknown, is connected to the trolley are:
•
the EARTH LEAKAGE CURRENT of the equipment could lead to
an excessive EARTH LEAKAGE CURRENT of the system, in
normal condition;
•
the ENCLOSURE LEAKAGE CURRENT of the equipment could
lead to an excessive ENCLOSURE LEAKAGE CURRENT of the
system, in single fault conditions (interruption of common protective
earth conductor).
Mechanical
•
Block the handle before removing the Videokeratoscope from the staff The arm
of the OR trolley base is weight-balanced for an average weight of the Scout VK.
Prior to remove the VK, you should block the spring mechanism within the arm by
the proper handle, or the staff can hurt your face.
•
Insert the pin of the
Slit Lamp Adaptor
into the Slit Lamp hole before tightening
the hand-wheel. Tightening of the knob with the pin only partly inserted can
permanently deform it.
Cod. 161204EN
3-3