Keratron
®
Scout Installation and Operating Manual
OPTIKON 2000
PARAMETER SPECIFICATION
ENVIRONMENTAL SPECS
Storage: .......................................
temp range -10°C to +60°C,
humidity 0-100% (cond.included)
atm. pressure 500 to 1060 hPa
Operation: ....................................
temp range 0°C to +40°C,
humidity 0-90% (non condensing)
atm. pressure 700 to 1060 hPa
ELECTRICAL SPECS
DOCKING BASE AND OR TROLLEY
Input voltage: ...............................
100/120/220/230-240 V A.C. select.
Frequency:................................... 50/60
Hz
Power consumption: ....................
60 VA
Line fuses: ...................................
220/240 Volt: 1 A T
100/120 Volt: 2 A T
KERATRON
®
SCOUT
Input voltage: ...............................
14-18 Vdc supplied by the Docking Base, by the Battery
or by the OR Trolley
Power consumption: ....................
28 VA
ACCURACY
Simulated K-Readings: ................ /- 0.25D on a normal cornea.
Focusing ...................................... BFS deviation: with Slit Lamp Adapter typical /0.15D
/- 0.25D handheld.To be checked by the
“repeatability check” function.
Misalignement.............................. Mean deviation on the map: +/- 0.1 with misalignement up
to 1mm.
EQUIPMENT CLASSIFICATION ACCORDING TO IEC 60601-1
Type of protection against electric shock :
Docking Base...............................
Class I
OR Trolley....................................
Class I
Keratron
®
Scout ...........................
Class II
Cod. 161204EN
5-3