U
T
is the a.c. mains voltage prior to application of the test level
Note 1:At 80 MHz and 800 MHz, the higher frequency range applies.
Note2:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To access the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Keratron
®
Scout is used exceeds the applicable RF
compliance level above, the Keratron
®
Scout should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Keratron
®
Scout unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
5.3.3 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND
MOBILE RF COMMUNICATION EQUIPMENT AND
KERATRON
®
SCOUT
Keratron
®
Scout is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of Keratron
®
Scout can help preventing electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and Keratron
®
Scout as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
(m)
Rated maximum
output power of
transmitter (W)
150 kHz to 80 MHz
d= 1.2
√
P
80 MHz to 800 MHz
d= 1.2
√
P
800 MHz to 2.5 GHz
d= 2.3
√
P
0.01
0.12
0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rates at maximum output power not listed above, the
recommended separation distance in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.