76
Basic definitions of parameters
may be viewed by touching the Fv/m section of
the Fv/Fm measuring screen.
Fv/Fm Test Setup Screen
Error messages
common to the test modes:
There are several common errors that can occur in the test modes.
If the battery voltage is 10.5 or less, the warning message “!! Battery Low !!” will appear
and a long beep will be sounded.
If the amount of in-band IR radiation is above set limits thereby saturating the pre-
detector the error message “!! IR to High !!” will be displayed.
“Fluorescent signal too low” appears if there is not enough signal getting to the detector.
This may be due to the fiber optic being too far from the sample, the intensity of the
modulated source being too low or the gain control set too low.
“FL > 3600 pos det pin out” appears if the combination of the modulated light intensity
and modulated light gain are set too high. To fix the problem go to Auto Gain on the
setup screen, put a leaf in the PAR Clip, and press Auto Gain. The modulation intensity
will adjust automatically.