72
One can also test for the proper saturation flash width by using the flash zoom and trying
various saturation flash widths. The diagram below explains how to evaluate the flash.
With the OS1p either the center or the right peak will provide optimal saturation pulse
settings.
Entering a note with a measurement
Fv/Fm measuring Screen 1p Test Data Collection Control Screen
A new feature included with the OS1p makes it possible to enter a note with any
measurement, or to require a note with every measurement. To enter a note, press the
“Log Setup: button. Next press “Enter a Note”. A touch keyboard will appear and one
can enter up to a 38 characters note with each measurement.
To require the note screen to appear with each measurement, press the “Add Notes to
sample” button and turn the feature ON. The keyboard screen will appear after the
measurement is complete. If the occasional note is needed, turn the feature off by
touching the “Add Notes to sample” button.