135
RLC Measuring Screen Step Set Up Screen Individual Step Set Up Screen
When a PAR Clip is not used “Default PAR”
When a PAR Clip is not attached to the OS1p, the user may select the approximate pre-
illumination intensity. From the RLC measuring screen, press the third window down
from the top on the left had side to get to the “Step Set Up Screen” shown above. Next,
press each step button that will be used in the RLC and enter a measured “Step Default
PAR Value”. When the “Step Default PAR Value” is pressed, a key board screen will
appear. “Default PAR” can be set from Zero to 2465
:
mols. OSI makes no claims about
the accuracy of “Default PAR” values. The actual value should be read with an
independent PAR meter at the leaf plane and leaf angle. When “Default PAR” is used, all
measurements of ETR appear using the setting selected. When a PAR clip is connected to
the OS1p, the “Default PAR” setting is automatically deactivated.