75
ZJ-SD
User’s Manual
S
e
ct
ion 5
DETAILE
D
S
E
TTINGS
Section 5
Linear Output
6.
Press the ENT key.
7.
Use the LEFT and RIGHT keys to display [L-
ADJ] on the main display.
The units for the monitor focus settings (mA or V) are
displayed on the sub-display.
8.
Press the ENT key.
The display changes to the display for setting the first
point (A).
The output current value is displayed on the main
display.
The correction value is displayed on the sub-display and
the leftmost digit flashes.
Setting the first point (A)
9.
Set the output current and correction values
for the first point.
Adjust the correction value on the sub-display so that the
ammeter reading and the output current shown on the
main display are the same.
The larger the correction value, the larger the output current.
The correction value can be set within the range -999 to 999.
To set a negative value, make the leftmost digit of the
sub-display flash and change the value.
If the correction value changes by 4, the linear output will
change by approx. 1.4 µA (0.7 mV).
The flashing digit, i.e., the digit to which a value can be set, changes as shown in the diagram
.
10.
Press the ENT key to confirm the setting.
This confirms the correction value for the first point.
The screen for setting the second point is displayed.
Moves from one digit to
another.
Changes the numeric value.
Digit shift on sub-display
Shifts one
digit at a time.
Switching from the sub-display to the main
display and vice versa
(from the leftmost/rightmost digit of the
sub-display)