Section 5
Linear Output
13.
Press the ENT key to confirm the setting.
This confirms the setting for the first point.
The screen for setting the second point is displayed.
Setting the second point (B)
14.
Use the same procedure as for the first point
to set the output current value and
corresponding measured value for the
second point.
15.
Press the ENT key to confirm the setting.
Confirming completion of monitor focus settings
The display indicates [OK] if monitor focus has been set
correctly.
If not, the display indicates [NG].
If [NG] is displayed, check the following points and
execute the monitor focus again.
• Is the measured value set on the sub-display within the
measuring range (with scaling and calculation settings
reflected if set)?
• Are the measured values of the first and second points
separated by at least 1% of the rated measuring range?
• Are the current (or voltage) values for the two points the
same?