2-2. Basic Measurement
PCG Messages
The PCG messages are explained below. When "OK" is not displayed, measurement accuracy
may be reduced. Take corrective action according to the content of the message.
Messages
Status
Corrective Action
OK
PCG is stable.
Measurement can begin.
Initializing
PCG initializing.
Have the patient remain quiet and wait briefly.
Out of
Range
The sensor has been touched, causing
noise.
Have the patient rest quietly and wait briefly.
Weak Signal
The PCG sensor has come off or the
patient's shirt or other clothing has
come under it.
Make sure the PCG sensor is in full contact with
the skin. If it is difficult to keep the sensor in full
contact, use the PCG sensor weight (refer to
page 46).
The PCG sensor pad is dry or dirty.
Replace with a new PCG sensor pad (refer to
page 45).
The wrong side of the PCG sensor is
attached.
Attach the PCG sensor correctly (refer to page 45).
A position where the first three or more levels of the meter are lit is optimum. Measurement
can be performed if the meter is only lit to the second or first level, but accuracy may be
reduced.
Re-Position
The position of the sensor is not
suitable.
Re-attach the PCG sensor in a different position
(refer to page 45).
For patients with cardiac murmur or respiratory noise, it may be difficult to clearly distinguish
the first and second heart sounds. While this may cause reduced accuracy, measurement is
still possible.
Note:
If "PCG" was set to "OFF" in the "SELECT MEASUREMENT SENSOR" settings when the
patient's information was entered (refer to page 34), PCG messages and the PCG level
indicator will not appear.