136
5-8. Specifications
Specifications of BP-203RPE
3
(Main unit)
General:
Name:
Non-invasive Vascular Screening Device
Model:
BP-203RPE
3
(HFA-RPE3-C3)
Type:
Non-invasive Vascular Screening Device
Dimension (W x H x D):
310 x 110 x 270 mm (excluding protrusions)
Weight:
Approx. 4.1 kg
Display part:
Method
8.4" TFT color LCD
Resolution
640 x 480 pixel
Safety standards:
EN/IEC60601-1
Medical electrical equipment-Part 1:
General requirements for safety
MDD class:
Classification Class
2
a
Protection class:
Class
1
Degree of protection:
NIBP
Type BF with defibrillator protection
ECG
Type CF with defibrillator protection
PCG
Type BF with defibrillator protection
Mode of operation:
Continuous
Other standards:
IEC62304
Medical device software-software life cycle processes
ISO14971
Medical devices-Application of risk management to medical devices
Environmental Conditions:
Power supply:
Rating
AC 100-240 V
Frequency
50/60 Hz
Power consumption
150 VA
Fuses
Quantity 2, 250 V, T2AH (Time-lag, High Breaking Capacity (HBC))
Operational temperature and humidity:
Temperature range
+10 - +40°C
Humidity range
15 - 85% (no condensation)
Atmospheric pressure
700 - 1060 hPa
Storage and transportation:
Temperature range
-20 - +60°C
Humidity range
10 - 95% (no condensation)
Atmospheric pressure
500 - 1060 hPa
Dust and water resistance:
Class IPX0
Reference: IEC60529:
Degrees of protection provided by enclosures (IP Code)
EMC:
Reference standard
EN/IEC60601-1-2
Medical electrical equipment-Part 1-2:
General requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements and tests
BP-203RPE3(A).fm 136 ページ
9年
月
日 月曜日 午前
時
4分