5-9. Guidance and Manufacturer's Declaration
The BP-203RPE
3
is intended for use in the electromagnetic environment specified below. The
customer or the user of the BP-203RPE
3
should assure that it is used in such an environment.
Warning
Use only the specified supplies for the cord, cuff, and other parts. Do not install other than
specified options. Otherwise and accident may result.
Electromagnetic emissions IEC60601-1-2
Emissions Test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The BP-203RPE
3
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The BP-203RPE
3
is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity IEC60601-1-2
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
70 %
U
T
(30 % dip in
U
T
)
for 25 cycle
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
70 %
U
T
(30 % dip in
U
T
)
for 25 cycle
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the BP-203RPE
3
requires
continued operation during power mains
interruptions, it is recommended that the
BP-203RPE
3
be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.