1-4. Safety Information
Do not attach the arm cuff for patient during medical treatment with intravenous drip or blood
transfusion. This may result incorrect diagnosis and treatment.
Do not use this device on a patient that cannot express pain. If cuff pressurization does not stop in a
specified time, internal hemorrhaging may occur in the upper arm or ankle.
This device cannot be used during MRI imaging. This may cause an accident.
Pay constant attention to patient's condition during measurement. Otherwise an accident may result.
Do not perform measurement without cleaning and sterilizing the device, cuffs and sensors after
measuring for patient with infection.
Pay extra attention to a patient with pacemaker during measurement. Rate meters may continue to
count the pacemaker rate during occurrence of cardiac arrest or some arrhythmias. Do not rely
entirely upon rate meter alarm. Refer to the pacemaker pulse rejection capability in this manual.
Installation
As this device conforms to the IEC60601-1-2 standard, it can be used in combination with another
medical device. While or after an electrosurgical knife or a device which generates noise being used
near this device however, make sure that it is properly functioning. Otherwise incorrect diagnosis
and treatment or device failure may result.
Do not use this device in the presence of a flammable gas such as a highly inflammable anesthetic,
or in a high-pressure oxygen chamber or tent. Fire and explosion may result.
Do not install or store the device in a location where water or chemicals may splash on the device.
Electric shock may result.
Do not connect any electric device, without the approval of IEC60601-1, or not fulfilling IEC60601-1-1.
For the use of a PC or a printer with this device, the connecting device should be approved
according to standards mentioned above. (This unit meets the restricted level of leakage current
required for medical devices and it does not include all the connected devices. Connecting to other
device is not allowed unless the total leakage current of such a combination is within the restricted
level.) Neglecting this caution, it could cause electric shock to the device.
Maintenance
After cleaning the device, dry it completely before turning on the power again.
Otherwise electric shock or current leakage may result.