137
5-8. Specifications
Non-invasive Blood Pressure (NIBP):
Measurement method:
Oscillometric method
Measurement technology:
Linear deflation
Pressure display range:
0, 10 - 300 mmHg
Pressure display accuracy:
Less than ±3 mmHg
NIBP Measurement range:
[Arm]
SYS
60 - 250 mmHg
MAP
40 - 235 mmHg
DIA
40 - 220 mmHg
Pulse rate
40 - 180 bpm
[Ankle]
SYS
40 - 250 mmHg
MAP
30 - 235 mmHg
DIA
25 - 220 mmHg
NIBP accuracy:
Mean error and standard deviation per
ANSI/AAMI SP-10: Manual, electronic, or automated
sphygmomanometers
Usable cuff size:
10 - 15 cm (width of bladder)
Numbers of cuff:
4 Right and Left Brachials, Right and Left Ankles
Reference standards:
EN/IEC60601-2-30
Medical electrical equipment Part 2-30:
Particular requirements for safety, including essential performance
of automatic cycling non-invasive blood pressure monitoring
equipment
EN1060-1
Non-invasive sphygmomanometers-Part 1:
General requirements
EN1060-3
Non-invasive sphygmomanometers-Part 3:
Supplementary requirements for electro-mechanical blood
pressure measuring systems
Pulse Wave Velocity (PWV):
Pulse Wave Velocities are measured by the heart sound and dicrotic notch of each cuff pulse.
R wave of ECG counts the heart rate and makes the noise elimination window for the heart sound.
BP-203RPE3(A).fm 137 ページ
9年
月
日 月曜日 午前
時
4分