DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
263
Figure 5
‑
198 The support ring’s signal clipped (
left
) and with adjusted gain (
right
)
To calibrate frequency 2
1.
Press the MAIN FILTER menu key (
), followed by
FREQ 2
(A key), and then
rotate the knob until
200 Hz
or less is displayed.
2.
Press the DISP menu key (
), followed by
CHANNEL
(B key), and then rotate
the knob until
FREQ 2
is displayed.
3.
Place the probe in a defect-free area of the calibration standard near the thru-wall
flaw, and then press the NULL foot switch.
4.
Scan the entire length of the tube, and press the FREEZE key (
) when done
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...