DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
203
Make sure that the guides are properly adjusted for both fastener rows. Failure to
properly center the probe will lead to a very difficult calibration and possibly an
unusable setup.
To set the initial NORTEC 600 configuration
1.
Connect the probe and the cable to the PROBE connector on the NORTEC 600.
2.
Press the ADV SETUP menu key (
) once, and select
APPL SELECT
(A key)
to open the application selection menu. Select
Dual Frequency Sub
‑
Surface
with
the knob, then press
to accept (see Figure 5-115 on page 203).
Figure 5
‑
115 The Dual Frequency Sub
‑
Surface application
3.
Press the MAIN FILTER menu key (
) four times, and then set the
LO PASS
(B key) to 50 Hz using the knob.
4.
Press the MAIN FILTER menu key (
) again, and then set the
PRB DRV
(A key) to
HIGH
.
5.
Press the DISP menu key (
) twice, and then set the
POSITION
(C key) to
BOT RGHT
.
IMPORTANT
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...