DMTA-10040-01EN, Rev. E, February 2018
Control Functions
85
SCAN RPM
(NORTEC 600S and NORTEC 600D models only)
The
SCAN RPM
function controls the scanner rotational speed (revolutions per
minute) when an optional rotating scanner is connected to the NORTEC 600S
(N600S) or NORTEC 600D (N600D) instrument models.
To adjust the
SCAN RPM
setting, press the MAIN FILTER menu key (
) two
times, followed by the E key, and then rotate the knob to the desired value.
4.3.3
Special Menu — MAIN FILTER Key
PRB DRV (probe drive)
The NORTEC 600 instrument has three levels of probe drive that can be selected:
LOW
,
MEDIUM
, and
HIGH
. The approximate peak-to-peak voltages are 2 V, 6 V,
and 12 V, respectively.
MEDIUM
probe drive (the default setting) is normally sufficient for most eddy
current testing. However,
HIGH
probe drive is desirable in the following cases:
a)
If gain is insufficient at the lower probe-drive settings.
b)
During testing of lower conductivity materials.
c)
For finding smaller flaws in the test material.
d)
For deeper penetration into the test material.
To adjust the probe-drive level, press the MAIN FILTER menu key (
) three
times, followed by the A key. With
PRB DRV
highlighted, rotate the knob to the
desired level.
PRB CONN (probe connection)
The NORTEC 600 instrument supports two types of probe connections: BNC and
16-pin LEMO. The probe input by default is set to 16-pin LEMO. If the BNC
connector is used, the connection input needs to be changed manually.
To adjust the probe-connection input, press the MAIN FILTER menu key (
)
three times, followed by the B key. With
PRB CONN
highlighted, rotate the knob
to select the desired connector:
LEMO
‑
16
or
BNC
.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...