DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
197
Figure 5
‑
107 Adjusting the H GAIN for frequency 1
14. Place the probe on position 1 again, and then press the A-LIFT NULL key (
).
15. While firmly pressing the probe at position 2 on the standard, observe the
MIX
signal point, and then press the MAIN FILTER menu key (
) and decrease the
H GAIN 2
(D key) until the
MIX
signal operating point does not extend more
than 3 divisions to the left of the null point (see Figure 5-108 on page 197).
Figure 5
‑
108 Adjusting the H GAIN 2 value