43
1.
Spray the
Disinfectant
on the chosen piece and let it act for 15 minutes.
2.
Rub the piece gently with the
Wipe
.
3.
If some halos are present on the piece (this may be the case of
Acrylic
), spray the
Disinfectant
again and
immediately remove it gently with the
Wipe
.
Note
►
Please contact Okolab if you need further information about which Disinfectant and Wipe to use.
Filters Replacement
If you have purchased also H201-T-FILTER-BOX, a module containing a filters kit composed by a prefilter
and a HEPA filter (High Efficiency Particulate Air Filter) has been added to the Heating Box, as described in
paragraph 8.2.1.3. When activating the Filter Monitoring, the counting of the usage days,
indicated with ‘Update
Time’, starts. The maximum usage time
for the Prefilter is 180 days, while for the Hepa Filter is 360 days. When
the usage time exceeds one or both these values, you should replace them in order to avoid contaminations of the
sample.
Before opening the H201-FILTER-BOX, make sure that the Heating Box is unplugged from an outlet.
Before opening the H201-FILTER-BOX to change the Prefilter or the HEPA Filter wear personal
protective equipment, breathing mask and gloves.
Follow the indications below for the filters replacement:
1.
Open the two soft draw latches on the sides of the H201-FILTER-BOX, see A in Image 1 Figure 69
2.
Slip off the frontal panel of the H201-FILTER-BOX and pull out the prefilter and HEPA filter. See Image 2 and
3 of Figure 69.
3.
Replace one or both the filters that you need to replace with the new filters.
4.
Reassembly the H201-FILTER-BOX.