9
Calibration
Cage-Calibration
Carefully read the instructions given in this paragraph before starting the Cage Calibration.
Note
►
It is important to monitor the room temperature during the calibration, as well as reducing air drafts
and forced convection. If possible divert any draft coming from the air conditioning system in the room, from
the microscope incubation system.
The goal of this procedure is to adjust the temperature offset of the Cage temperature hence maintaining
the sample at the desired temperature (e.g. 37°C) during imaging. Before starting the Cage -Calibration:
1.
Plug the green Fine Gauge Thermocouple in the port labeled
“T sensor”
(see Figure 5 pointer 8) located in
the rear panel of the Temperature Control Unit