26
a
b
Figure 35. Touch screen settings. Date and Time (a
–
b)
8.2.2.5
Alarms
Press on the
Alarms
icon
as shown in Figure 36 a. Then press on the
Temperature
icon
(see
a
b
Figure 36. How to enter in Alarms page (a
–
b).
Tip
►
Flag “Buzzer” if you want the Alarm to be acoustical as well rather than just being displayed.
The page shown in Figure 37
will open and you can choose the “
Temp Deviation
” and “
Temp Time
”.
Tip
►
”Temp Deviation”
is the value defining the
allowed tolerance from the set point. “Temp Time” is the
time for which the temperature may remain outside the allowed tolerance before the controller triggers in
alarm.
Figure 37. How to set the Deviation and Time Alarm